Pulse Width
Pulse
Rise
Time
Slow
Function Group
MODULATION
Menu Map
4
Description
This
softkey
lets you set the swept signal generator to apply a slow
rise pulse filter to both internal and external pulse waveforms. This
results in pulses having approximately 2 ps rise/fall times. An
asterisk next to the key label indicates that this function is active.
Programming Codes
SCPI: PULM:SLEW
<num>[time
suffix]MAXimum
Analyzer: NONE
See Also
Pulse Menu
Pulse
Width
Function Group
Menu Map
Description
Programming Codes
See Also
(MOD)
4
This softkey lets you set a value for the internal pulse generator’s
pulse width. The range of acceptable values is from 1
11s
to 65.5 ms.
The factory preset value is 1 ms. When this feature is active, its
current value is displayed in the active entry area.
SCPI: PULSe:INTernal:WIDTh <num>[time suffix] or
MAXimumIMINimum
Analyzer: NONE
[E),
(l\/roD],
Pulse Menu
Operating and Programming Reference P-23