EasyManua.ls Logo

HP 8360 B Series - Programming Language Comparison

HP 8360 B Series
496 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 3-9. Programming Language Comparison
Description
Network Analyzer
Language
SCPI Language
ALC
Leveling mode, external
A2
POW:ALC:SOUR DIOD;
:POW:ATT:AUTO OFF
Leveling mode, internal
Leveling mode, mm module
Al
SHA2
POW:ALC INT
POW:ALC:SOUR MMH;
:POW:ATT:AUTO OFF
Leveling mode, power meter
A3
POW:ALC:SOUR PMET;
:POW:ATT:AUTO OFF
Enable normal ALC
operation
Al]A2]A3(SHA2
POW:ALC:STAT ON
Disable ALC and control
modulator drive directly
Set output power, then
disable ALC
SHA3
POW:ALC:STAT OFF
SHAl
POW:SEAR ON
Uncouple attenuator,
control ALC independently
Frequency
SHPS
<num>DB
POW:ATT:AUTO OFF;
:POW
<num>[DBM]
Set CW frequency
CW <num>freq-suffix
FREQ:CW <num>[freq-suffix]
;MODE
CW
Set start frequency
FA <num>freq-suffix
FREQ:STAR
<num>[freqsuffix]
;MODE
SWE
Set stop frequency
FB <num>freq-suffix
FREQ:STOP
<num>[freqsuffix]
;MODE
SWE
Set center frequency
CF <num>freq-suffix
FREQ:CENT
<num>[freqsuffix]
;MODE
SWE
Set frequency span
DF <num>freq-suffix
FREQ:SPAN <num>[freq-suffix]
;MODE
SWE
Set swept mode step size
Set
CW mode step size
Enable frequency
offset function
Enable frequency
nultiplier function
SHCF <num>freq-suffix
SHCW <num>freq-suffix
SHFB <num>freq-suffix
SHFA
<num>
FREQ:STEP
<num>[freq-suffix]
FREQ:STEP <num>[freq-suffix]
FREQ:OFFS
<num>[freq-suffix]
;OFFS:STAT
ON
FREQ:MULT
<num>;MULT:STAT
ON
Keep
multiplication factor on
instrument on/off or preset
Multiplication factor=1 on
instrument on/off or preset
SHAL
(Refer to user defined preset)
SHIP
(Refer to user defined preset)
Zoom function
SHST
Installation 3-25

Table of Contents

Related product manuals