Doubler
Amp
Mode
On
Function Group
Menu Map
Description
Programming Codes
See Also
POWER
5
This
softkey
is applicable to instrument models with a doubler
installed. The doubler has an integral amplifier whose operation is
controlled by the instrument firmware. This softkey turns off the
automatic mode of operation and turns on the amplifier so that it is
always used. This is an unspecified mode of operation since it can
cause increased harmonics and degraded dynamic range at some
frequencies. This
softkey
has no effect on instruments without a
doubler.
An asterisk next to the key label indicates that this feature is active.
SCPI:
POWer:AMPLifier:STATE ON]OFF]O]l
POWer:AMPLifier:STATE?
Analyzer: NONE
Dblr Amp Menu
Dwell
Coupled
Function Group
Menu Map
Description
Programming Codes
See Also
FREQUENCY
2
This softkey lets you couple the dwell time for points in the stepped
frequency sweep mode to the ramp sweep mode sweep time. The
equation to determine the dwell time in the dwell coupled mode is as
follows:
Coupled Dwell Time = (sweep time)
+
(number of step points)
An asterisk next to the key label indicates that this feature is active.
SCPI: SWEep[:FREQ
uency]:DWELl:AUTO
ON]1
Analyzer: NONE
Step
Swp
Menu
D-10 Operating and Programming Reference