P
Peak
RF
Always
Function Group
POWER, USER CAL
Menu Map
5,9
Description
This
softkey
appears in two locations: the POWER Tracking Menu
and the USER CAL Tracking Menu. The operation is the same in
both locations.
This
softkey
causes the swept signal generator, when in CW or
manual-sweep output mode, to align the output filter (SYTM) so
that its
passband
is centered on the RF output. Peaking is used to
obtain both the maximum available power and spectral purity, and
the best pulse, FM envelopes, at a given frequency. This peaking
occurs each time the frequency is changed, or every seven minutes.
An asterisk next to the key label indicates this function is active.
Programming Codes
SCPI: CALibration:PEAKing:AUTO
ONlOFFlllO
Analyzer:
RPl
function on, RPO function off.
See Also
Auto Track, Peak RF Once, Tracking Menu
“Optimizing Swept Signal Generator Performance” in Chapter
1
Operating and Programming Reference P-l