Figure l-19. Creating a User
Flatness Array
POWER
NETER
-00cl
P
.
6
.El
Q
For this example, refer to menu map 5, POWER.
1. The equipment setup shown in Figure 1-19 assumes that if your
setup has an external leveling configuration, the steps necessary
to correctly level have been followed. If you have questions about
external leveling, refer to “Externally Leveling the Swept Signal
Generator”.
Setup Power Meter
2. Zero and calibrate the power meter/sensor.
3. Connect the power sensor to the point where flatness corrected
power is desired.
Setup Swept Signal Generator Parameters
4. On the swept signal generator, press (PRESET).
5. [POWER LEVEL)
@
(dB0).
This sets the test port power to
+5
dBm
(PO
maX
-
Ppath
loss).
Create A Frequency List
6. On the swept signal generator, press FREQUENCY (MENU).
7.
Select List
Menu
Enter List Freq
a@&.
This enters
5 GHz as the first frequency in the list array. Entering a
frequency automatically sets the offset to 0 dB and the dwell to
10 ms.
8. Enter 18, 13, 11, and 20 GHz to complete this example array.
Getting Started Advanced 1-37