HT2000H_en_n.doc Page 18 of 167
ENVIRONMENT CONDITIONS
Room temperature limits 15°C - 35°C
Ambient humidity limits 5% - 80% (not-condensable)
Maximum altitude 2000m
SOUND PRESSURE LEVEL
Maximum measured level 60dBA (below the limits of 85dBA defined
by the regulations in force)
SAFETY INFORMATION
The autosampler is classified as shown below:
Pollution degree 2
Overvoltage category II
Devices for use indoors
TRAY AND SYRINGE FEATURES
Tray capacity 42 vial positions (HT2000H and HT2000HT)
14 vial positions (HT2100H)
Syringe volume 1-2.5-5ml
Maximum pressure in the injector 5 bar
(2.5ml syringe)
ANALYTICAL PARAMETERS
Conditioning
HT2000H
Oven positions 6
Oven Temperature off, from 40°C1 to 170°C
Incubation time from 0 to 999min step 1min
Shaking method orbital
Shaker speed from very low to very high
Shaking cycles on/off, from 0 to 9.9min
HT2100H
Oven positions 1
Oven Temperature off, from 40°C2 to 150°C
Incubation time from 0 to 999min step 1min