The calibration used for calculating the result was obtained with at least one calibrator
excluded.
For the Clot Detection option
A clot was detected when pipetting on the sample.
All the calculated results must be verified.
Result higher than the higher normal value.
Spectrophotometry
The absorbances measured while monitoring the reaction are not strictly increasing or
decreasing (non-monotony of the reaction).
* * * * * is displayed instead of the result.
Result lower than the lower normal value.
Spectrophotometry
The result was calculated from absorbencies measured with an unstable lamp.
The assay is no longer being requested by analyser.
Luminescence
This message is applicable to semi-quantitative assays.
The luminescence signal measured is below the minimal acceptable signal.
In this case, the sample must be tested again.
* * * * * is displayed instead of the result.
Luminescence
Error when recovering the measurements.
Spectrophotometry
Error when recovering the measurements, or the result can not be calculated because the
number of measurements insufficient.
Luminescence
The calibration used for calculating the result was obtained with at least one relative light
unit (RLU) measurement excluded.
It has not been possible to calculate the result because the calibration is not valid.
* * * * * is displayed instead of the result.
The result calculated is below the lower limit of the field of measurement.
«< Low limit of the sample measurement range» is displayed instead of the result.
Luminescence
This message is applicable to semi-quantitative assays.
The result calculated is below the lower limit of the field of measurement and forced to 0:
the sample result can be reported as Negative.