The result calculated is above the upper limit of the field of measurement. The sample will
be automatically diluted if a dilution rate is defined in the analytical configuration.
«> High limit of the sample measurement range» is displayed instead of the result.
If the message OMR+ accompanies the message “RED”, the result is still higher than the
upper limit of the field of measurement after dilution.
In this case, the sample must be re-diluted to a level higher than that programmed in
the analytical configuration, then tested again.
Luminescence
The result cannot be calculated as the measurement is outside the scope of the
measurements defined by the master curve. * * * * * is displayed instead of the result.
Luminescence
Problem with luminometer stability.
Luminescence
The calibration used for calculating the result was obtained with one of the replicates for
calibrator excluded.
Immunoassay
Problem with agitation of magnetic particles on the reagent rack.
The result was recalculated after modifying the calibration.
The result was calculated after a dilution. The value given takes into account the dilution
factor.
Result obtained for a reflex test.
Storage temperature of reagents outside limits during the assay.
Spectrophotometry
Detection of enzymatic hyperactivity. The change in absorbance per minute (∆A/min)
calculated between the injection of the sample and the first measurement used in the
calculation exceeds the programmed limit.
* * * * * is displayed instead of the result.
The sample will be automatically diluted if a dilution rate is defined in the analytical
configuration.
If the SUB message accompanies the message “RED”, enzymatic hyperactivity is still
detected after dilution.
In this case, the sample must be re-diluted to a level higher than that programmed in
the analytical configuration, then tested again.
Luminescence
At least two measurements of the integrated signal are out of the scope of the
luminometer, * * * * * is displayed instead of the result.
Luminescence (if thermal compensation activated)
Ambient temperature lower than the acceptable limit.
Luminescence (if thermal compensation activated)
Ambient temperature higher than the acceptable limit.
One of the Westgard rules applied for this analyte has been violated.