The pipetting arm will move to the rinsing well before returning to the new adjusted position :
•
Check the probe position, then remove the lid support tool and close the lid (see page 36).
•
Click on .
The pipetting arm will move to the rinsing well.
A message confirming completion of the procedure is displayed :
•
Click on to validate this message.
A window containing the settings values is displayed :
•
Click on .
The main position adjustment window is displayed.
•
Click on to exit this menu
Maintenance
5-4- Analyser Interventions (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-4
135
Software version V 14