The pipetting arm will move to the rinsing well before returning to the new adjusted position :
Check the probe position, then remove the lid support tool and close the lid (see page 36).
Click on .
The pipetting arm will move to the rinsing well.
A message confirming completion of the procedure is displayed :
Click on to validate this message.
A window containing the settings values is displayed :
Click on .
The main position adjustment window is displayed.
Click on to exit this menu
Maintenance
5-4- Analyser Interventions (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-4