Click on .
The pipetting arm moves to its home position and the lid is unlocked.
The message displays:
•
Open the lid and insert the lid support tool (see pages 35 and 36).
•
Insert firmly the activation key into place.
WARNING:
USE OF THE ACTIVATION KEY
The activation key allows a particular use of the analyser, without user’s protection
against the pipetting arm movements. The pipetting arm will move with the lid
opened.
Contact with the moving pipetting arm may cause severe injuries.
The use of the activation key is strictly restricted to this adjustment procedure.
•
When the activation key is detected, the message displays:
Maintenance
5-4- Analyser Interventions (continued)
IDS-iSYS User Manual - Revision M1!
Maintenance 5-4
137
Software version V 14