For systems using DSORB solution, a maintenance procedure for the four washers will take place
automatically during the SHUT DOWN procedure. Each shutdown procedure will use the following products:
Description
Quantity
IDS-iSYS WASH S
24 mL
IDS-iSYS D-SORB
3 mL
IDS-iSYS Cuvettes
4 cuvettes
Remove all samples.
Remove reagent racks and store in accordance with IFU recommendations.
Carry out daily maintenance, then if necessary, weekly and monthly maintenance.
For systems using AP Substrate, from the MONTHLY MAINTENANCE menu, perform the « Rinsing of the AP
Substrate tubing» maintenance twice to rinse the fluidic circuit following the below steps:
Replace the AP Substrate bottle by a 500 mL bottle containing de-ionised water to mimic the AP
Substrate.
Click on the corresponding button in front of this item in the maintenance list.
Click on .When the first step is completed, a message is displayed.
Keep the distilled water bottle in the AP Substrate position, continue the maintenance by clicking on
.
Repeat this procedure a second time.
After the two rinsing procedures are completed, perform the « Rinsing of the AP Substrate tubing»
maintenance for a third time to compete the rinse and to empty the fluidic circuit:
Keep the distilled water bottle in the AP Substrate position.
Click on the corresponding button in front of this item in the maintenance list.
Click on .When the first step is completed, a message is displayed.
Replace the bottle containing distilled water with an empty bottle. Continue the maintenance by
clicking on .
Empty the liquid waste
Empty the solid waste.
Waste must be disposed of in accordance with current local regulations (see APPENDIX I: Waste disposal,
page A2).
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
Waste which contains, or has been in contact with, biological specimens must be
considered to pose a potential risk of infection.
Wear disposable gloves when handling waste.
The waste must be disposed of in accordance with current local regulations.
Click on .
Click on the liquid waste icon.
Then validate the message confirming the liquid waste emptying.
Appendices