Click on the solid waste icon.
Then validate the message confirming the solid waste emptying.
•
Check the current volume of each ancillary reagent is sufficient to cover the stoppage period.
If necessary, install a new ancillary reagent.
•
Exit the menu.
•
In SETUP menu, select the SETTINGS LOCAL SYSTEM menu:
•
Program a daily automatic wake-up, without exception days .
•
Program a daily automatic shut down, without exception days . One hour time delay between wake-up and
shut-down is required.
•
If necessary, unselect the automatic QC profile option.
•
Click on to save the modification.
•
From the Session menu, select SHUT DOWN to put the analyser into standby mode.
For systems using DSORB solution, a maintenance procedure for the four washers will take
place automatically during the SHUT DOWN procedure.