A6-2-Between 15 and 30 days
The analyser can be completely stopped after a decontamination of the IDS-iSYS TRIGGER B fluidic circuit
followed by the emptying of the two TRIGGER and of the D-SORB fluidic circuits. The TRIGGER tubing
cleaning maintenance, included in the weekly maintenance, will be used to carry out this procedure.
Remove all samples.
Remove reagent racks and store in accordance with IFU recommendations.
Carry out daily maintenance.
Carry out the general steps of the weekly maintenance.
Perform the «Rinsing of the Trigger A and Trigger B fluid lines» maintenance a first time for rinsing the fluidic
circuits:
Replace the two Trigger bottles by two 250 mL bottles containing distilled water.
When the first step of the maintenance is completed, continue the maintenance with the two bottles
containing distilled water.
When the rinsing is over, perform the «Rinsing of the Trigger A and Trigger B fluid lines» maintenance a
second time for decontamination of the fluidic circuits:
Keep the distilled water bottle in the TRIGGER A position. Install the TRIGGER A bottle at the
TRIGGER B position.
When the first step of the maintenance is completed, continue the maintenance with the same bottles
installed.
When the decontamination is over, perform the «Rinsing of the Trigger A and Trigger B fluid lines»
maintenance a third time for rinsing and emptying the fluidic circuits:
Replace the two Trigger bottles by two 250 mL bottles containing distilled water.
When the first step of the maintenance is completed, replace the two bottles containing distilled water
with empty bottles. Continue the maintenance.
Carry out the general steps of the monthly maintenance.
For systems using AP Substrate, perform the « Rinsing of the AP Substrate tubing» maintenance twice to
rinse the fluidic circuit following the below steps:
Replace the AP Substrate bottle by a 500 mL bottle containing de-ionised water to mimic the AP
Substrate.
Click on the corresponding button in front of this item in the maintenance list.
Click on .When the first step is completed, a message is displayed.
Keep the distilled water bottle in the AP Substrate position, continue the maintenance by clicking on
.
Repeat this procedure a second time.
After the two rinsing procedures are completed, perform the « Rinsing of the AP Substrate tubing»
maintenance for a third time to complete the rinse and to empty the fluidic circuit:
Keep the distilled water bottle in the AP Substrate position.
Click on the corresponding button in front of this item in the maintenance list.
Click on .When the first step is completed, a message is displayed.
Replace the bottle containing distilled water with an empty bottle. Continue the maintenance by
clicking on .
Remove the bottle in the AP Substrate position.
Appendices
IDS-iSYS User Manual - Revision M1!
Appendix A-6
A 11
Software version V 14