For rinsing the D-SORB fluidic circuit, replace the D-SORB bottle with a bottle containing de-ionised water.
Identify the position by using the barcode label of the D-SORB bottle that just has been removed from the
system.
Request three times the priming SPECIFIC DECONTAMINATING SOLUTION, from the menu
For emptying the D-SORB fluidic circuit, remove the D-SORB level sensor from the bottle and wipe it.
Invert the D-SORB level sensor and keep it in the air.
Identify the position by using the barcode label of the D-SORB bottle that just has been removed from the
system.
Request three times the priming SPECIFIC DECONTAMINATING SOLUTION, from the menu
Empty the liquid waste
Empty the solid waste.
Waste must be disposed of in accordance with current local regulations (see APPENDIX I: Waste disposal,
page A2).
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
Waste which contains, or has been in contact with, biological specimens must be
considered to pose a potential risk of infection.
Wear disposable gloves when handling waste.
The waste must be disposed of in accordance with current local regulations.
Click on .
Click on the liquid waste icon.
Then validate the message confirming the liquid waste emptying.
Click on the solid waste icon.
Then validate the message confirming the solid waste emptying.
Check the current volume of each ancillary reagent is adequate with the stop period.
From the File menu, click on EXIT, the software will close.
Then turn o the computer following the procedure for shutting down WINDOWS.
Finally, press the switch located on the left-hand side of the analyser into position ‘0’.