Information Application Buttons
Provides access to information regarding the analyser status
including :
•
General status of the apparatus (initialised, standing etc.).
•
Status of the various modules.
•
Allowed/forbidden assays depending on the field.
•
Temperatures : reagent compartment, carousel, ambient.
•
Status of reagents.
•
Status of reagent drawer, samples drawer, lid.
The general status is indicated by a colour code:
•
Green = Operational.
•
Orange = Caution: one of the elements is outside the optimal
operating conditions
(for example, temperature).
•
Red = Error/fault.
•
Yellow = Analyser not initialised (standing).
Provides information covering the workload requested including:
•
Missing reagents required to perform queued assays.
•
Expired reagents.
•
Unavailable tests.
•
Tes ts sto ppe d du r ing th e c y cle d u e t o ca lib rat i on fa i lure s,
Westgard rules violation etc.
•
Automatic requests for calibrations and controls generated by the
system.
If messages affecting the current workload appear, the button is
displayed in orange with the number of warning messages
indicated.
If automatic requests have been generated, a flashing icon
will appear.
Provides access to the alarms generated by the system:
•
Module errors.
•
Lack of reagents.
An error message is indicated in the ALARM button which is
displayed in red and contains the number of errors identified.
User Interface Software
2-2- Main Screen (continued)
IDS-iSYS User Manual - Revision M1!
User Interface Software 2-2
21
Software version V 14