The list of calibrations for each reagent lot is shown according to the selected display option:
Displays the last calibration for each on-board lot of reagent.
Displays all the calibrations performed in the last 60 days.
Displays the calibrations of all unexpired reagent lots for all
analytes of the private library.
Only the last calibration is diplayed for each lot of reagent.
Comment: The display option selected is kept after closing the window.
Programming a Calibration of a New Reagent Lot
When calibrating a new Immunoassay reagent lot, the calibrators associated with this lot must
be placed into the sample rack before programming the request.
•
Click on alongside the assay to be calibrated.
A message confirming the request displays:
•
Click on .
A tick is displayed in the calibration column and in the appropriate QC column(s).
Comment: The request can be cancelled by clicking on .
If more than one reagent lot is on-board, when is clicked on a new window appears (see below).
Use
3-5- Programming Calibrations and Controls (continued)
IDS-iSYS User Manual - Revision M1!
Use 3-5
56
Software version V 14