Select the new lot to be calibrated, and then click on .
A message confirming the request displays:
Click on .
A tick is displayed in the calibration column and in the appropriate QC column(s).
Comment: The request can be cancelled by clicking on .
Programming a New Calibration for a Reagent
The resulting screen (see below) shows a list of all tests in the personal library. It also includes the date and
time of the last calibration for the on-board analytes.
Click on the box in the column in front of the lot to be calibrated.
A message confirming the request displays:
Click on .
A tick is displayed in the calibration column and in the appropriate QC column(s).
Comment: The request can be cancelled by clicking on .
Use
3-5- Programming Calibrations and Controls (continued)
IDS-iSYS User Manual - Revision M1!
Use 3-5
57
Software version V 14