Excluding one of the replicates
For Immunoassay, individual outlying calibrator or control replicates can be excluded when the CV is above the
acceptable limit.
Click on to display the measurements (RLU or mAbs.).
To exclude one of the measurements, click on the corresponding red tick.
The calibration and the control(s) are calculated with the new curve equation.
When points have been excluded, the icon is displayed with the calibration date.
When the recalculated calibration is valid, with the control(s) within acceptable limits, the message OK is
displayed.
Click on to save the modification.
If the calibration is not validated, the button is displayed.
In this case, request a new calibration after having replaced failed calibrator/control.