6–13
Transpector CPM Operating Manual
Response Time Mode
12 Press  . The display will show the current Response Time. The default is 
20, at which the display color will change to red. 
13 Change the setting by pressing   or  . 
14 Press   to return to Measure Mode.
Upper And Lower Pressure Limits
15 Press  . The current lower limit is displayed. The display will alternately 
flash n_1 and a number representing the lower pressure limit. 
16 Press   to change the lower limit to 58 psi. 
16a The least significant digit will flash and can be changed using or . 
16b Press   to set the digit to the desired number.
16c The next digit will flash. 
16d Change the digit using either   or  , then press  . 
16e Continue this process until 58.0 is achieved. 
16f Press  . The display will flash 58.0 with a vertical bar showing to the 
left of the 58.0. 
17 Press  . The current upper limit is displayed. The display will alternately 
flash between n_2 and some number representing the upper pressure limit. 
Follow the same procedure (refer to step 16) to set the upper limit to 100 psi.