Multitron - Incubation Shaker
Operation
Page 82 of 146 25 February 2019
Carbon dioxide concentration (CO
2
)
The parameter carbon dioxide concentration (CO
2
long to the standard scope of delivery and therefore it is only avail-
able if the according option is installed.
When entering the setpoint value for the carbon dioxide concentra-
tion it is shown in percent with an accuracy of one decimal place
the parameter is not installed, the display remains empty.
Light intensity (Light)
The parameter light intensity (Light) does not belong to the stand-
ard scope of delivery and therefore it is only available if the accord-
ing option is installed.
When entering the setpoint value for the light intensity it is shown
in percent. The light intensity can be set from 1 up to 100 %. The
light intensity is linearly proportional to a value of 5 to
200 µmol m
-2
s
-1
.
If the parameter is not installed, the display remains empty.
Rotation speed (RPM)
The parameter rotation speed (RPM) belongs to the standard
scope of delivery.
When entering the setpoint value for the rotation speed it is shown
in rpm (min
-1
) with an accuracy of one whole number.
The highest rotation speed allowed depends on the throw, the
position of the equipment in a stack and the load on the tray