Multitron - Incubation Shaker
Operation
25 February 2019 Page 83 of 146
Humidity (Humid)
The parameter humidity (Humid) does not belong to the standard
scope of delivery and therefore it is only available if the according
option is installed.
When entering the setpoint value for the relative humidity it is
shown in percent with an accuracy of one decimal place.
If the parameter is not installed, the display remains empty.
Parameter-specific display elements
The sign Setpoint lights up, when by tapping the SET key on the
upper or lower operating panel the setpoint values for the parame-
ters can be entered.
The sign High or Low lights up, if the current value of a parameter
is above respectively below its setpoint value.
After a certain time span or when a critical value is reached an
alarm will be set off, because most probably an error or defect hin-
ders the equipment to reach or maintain the setpoint value in ques-
tion.
The possible setpoints for the parameters rotation speed and tem-
perature can be limited in the settings menu. If an attempt is made
to enter a higher value, the message bloc appears.