SCPI Command
[SOURce[1|2]:]SUM:SOURce {INTernal|CH1|CH2}
Frequency Sweep
In frequency sweep mode, the instrument moves from the start frequency to the stop frequency at a specified sweep
rate. You can sweep up or down in frequency, with either linear or logarithmic spacing. You can also configure the
instrument to output one sweep from start frequency to stop frequency by applying an external or manual trigger.
The instrument can sweep sine, square, pulse, ramp, triangle, or arbitrary waveforms (PRBS, noise, and DC are not
allowed).
You can specify a hold time, during which the sweep remains at the stop frequency, and a return time, during which
the frequency changes linearly from the stop frequency to the start frequency.
For more information, see Frequency Sweep.
To Select Sweep
The instrument will not allow sweep or list mode to be enabled at the same time that burst or any modulation mode
is enabled. When you enable sweep, the burst or modulation mode is turned off.
To avoid multiple waveform changes, enable the sweep mode after configuring the other parameters.
Front Panel Operations
Press [Sweep] > Sweep ON | OFF> Sweep ON | OFF.
SCPI Command
[SOURce[1|2]:]FREQuency:MODE SWEEP
[SOURce[1|2]:]SWEep:STATe {ON|1|OFF|0}
122
Keysight EDU33210 Series User's Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EDU33210 Series and is the answer not in the manual?

Keysight EDU33210 Series Specifications

General IconGeneral
BrandKeysight
ModelEDU33210 Series
CategoryPortable Generator
LanguageEnglish

Related product manuals