EasyManua.ls Logo

Keysight EDU33210 Series - Marker Frequency

Keysight EDU33210 Series
503 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SCPI Command
[SOURce[1|2]:]SWEep:HTIMe {<hold_time>|MINimum|MAXimum|DEFault}
[SOURce[1|2]:]SWEep:RTIMe {<return_time>|MINimum|MAXimum|DEFault}
Marker Frequency
If desired, you can set the frequency at which the signal on the front panel Sync Out connector goes to a logic low
during the sweep. The Sync signal always goes from low to high at the beginning of the sweep.
Marker frequency: 1 μHz to maximum frequency for the waveform. The default is 500 Hz.
When the sweep mode is enabled, the marker frequency must be between the specified start frequency and stop
frequency. If you attempt to set the marker frequency to a frequency not in this range, the instrument will set the
marker frequency equal to the start frequency or stop frequency (whichever is closer).
You cannot configure the marker frequency with the front panel menus unless the Sync source is the sweeping
channel.
Front Panel Operations
1. Press [Sweep] > Sweep ON | OFF > Sweep ON | OFF.
2. Press [Trigger] > Sync ON | OFF > Sync Setup.
3. Select Mode Marker.