In gated mode, burst count, burst period, and trigger source are ignored (used for triggered burst only). Manual
triggers ignored; no error generated.
In gated mode, you can specify polarity of signal on the front panel Ext Trig connector ([SOURce
[1|2]:]BURSt:GATE:POLarity {NORMal|INVerted}). Default is NORMal (true-high).
Front Panel Operations
Press [Burst] > N Cycle Gated or N Cycle Gated.
SCPI Command
[SOURce[1|2]:]BURSt:MODE {TRIGgered|GATed}
Waveform Frequency
You can specify the signal frequency during the burst in triggered and external gated modes. In the triggered mode,
the number of cycles specified by the burst count is output at the waveform frequency. In the external gated mode,
the waveform frequency is output when the external gate signal is true. This differs from the "burst period," which
specifies interval between bursts (triggered mode only).
Waveform frequency: 1 μHz to maximum frequency of the waveform. The default value is 1 kHz. (For an internally
triggered burst waveform, the minimum frequency is 126 μHz.)
Front Panel Operations
Press [Parameter] > Frequency. Use the numeric keypad or the knob and arrows to set a desired value. If you use the
keypad, select a prefix unit to finish.
134
Keysight EDU33210 Series User's Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EDU33210 Series and is the answer not in the manual?

Keysight EDU33210 Series Specifications

General IconGeneral
BrandKeysight
ModelEDU33210 Series
CategoryPortable Generator
LanguageEnglish

Related product manuals