2. Press Channel [On/Off] to produce the DC output.
SCPI Command
[SOURce[1|2]:]FUNCtion <function>
The APPLy command configures a waveform with one command.
Output Frequency
The output frequency range depends on the function, model, and output voltage, as shown here. The default
frequency is 1 kHz for all functions, and the minimum frequencies are shown in the table below.
Function Minimum Frequency
Sine 1 µHz
Square 1 μHz
Ramp/Triangle 1 μHz
Pulse 1 μHz
PRBS 1 mbps
Arbitrary 1 μSa/s
– Frequency limitations: Changing functions may change the frequency to meet the new function's frequency limits.
Arbitrary waveforms retain their last frequency setting.
– Burst limitation: For internally-triggered bursts, the minimum frequency is 126 μHz.
– Duty cycle limitations: For Square and Pulse, the Duty Cycle is limited by the 16-ns minimum pulse width
specification. For example, at 1 kHz, Duty Cycle may be set as low as 0.01%, because that would result in a pulse
width of 100 ns. At 1 MHz, the minimum Duty Cycle is 1.6%, and at 10 MHz it is 16%. Changing to a frequency that
cannot produce the current duty cycle will adjust the duty cycle to meet the minimum pulse width specification.
The minimum pulse width is 16 ns.
82