2 Applies to sample clock, not whole waveform
Frequency Limitations: Changing functions may change the frequency to meet the new function's frequency limits.
Amplitude Limitations: When the output units are Vrms or dBm, changing functions may lower the amplitude to the
maximum for the new function due to variation in waveform shapes. For example, a 5 Vrms square wave (into 50 Ω)
changed to a sine will decrease to 3.536 Vrms (sine’s upper limit).
Amplitude and offset cannot combine to exceed the instrument’s capability. The one you set last may be changed to
stay within limits.
You may protect a device-under-test (DUT) by specifying upper and lower output voltage limits.
Front Panel Operations
To turn on an output: Press Channel [On/Off] for your desired channel.
To select another waveform: Press [Waveform].
For example, to specify a DC signal:
1. Press [Waveform] > MORE 1 / 2 > DC > Offset.
Use the numeric keypad or the knob and arrows to set a desired value. If you use the keypad, select a unit prefix to
finish.