Front Panel Operations
1. Select Pulse waveform: Press [Waveform] > Pulse.
2. Select period instead of frequency: Press [Units] > Frequency Periodic > Frequency Periodic.
3. Set the period: Press [Parameter] > Period. Use the numeric keypad or the knob and arrows to set a desired value. If
you use the keypad, select a unit prefix to finish.
SCPI Command
[SOURce[1|2]:]FUNCtion:PULSe:PERiod {<seconds>|MINimum|MAXimum|DEFault}
Pulse Width
Pulse width is the time from the 50% threshold of a pulse's rising edge to the 50% threshold of the next falling edge.
Pulse width: up to 1,000,000 s (see restrictions below). The default pulse width is 100 μs. The minimum pulse
width is 16 ns.
The specified pulse width must also be less than the difference between the period and the minimum pulse width.
The instrument will adjust the pulse width to accommodate the specified period.
Front Panel Operations
Press [Waveform] > Pulse > Pulse Width. Use the numeric keypad or the knob and arrows to set a desired value. If
you use the keypad, select a unit prefix to finish.
96
Keysight EDU33210 Series User's Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EDU33210 Series and is the answer not in the manual?

Keysight EDU33210 Series Specifications

General IconGeneral
BrandKeysight
ModelEDU33210 Series
CategoryPortable Generator
LanguageEnglish

Related product manuals