Select ViewList to view the list parameters. You can edit (Edit Freq) the frequency value in the sweep list, add (Add
Freq) or delete (Delete Freq) a frequency value, and reorder the sweep list (Reorder List).
If you have a external USB flash drive connected, press Save List to save the sweep list to the external USB flash
drive.
To retrieve a previously saved sweep list from the connected external USB flash drive, press Select List.
SCPI Command
[SOURcd[1|2]:]FREQuency:MODE LIST
[SOURce[1|2]:]LIST:FREQuency <freq1>[, <freq2>, etc.]
Progress through list is controlled by the trigger system. If trigger source is internal or immediate, the dwell time
setting (LIST:DWELl) determines time spent at each frequency. For any other trigger source, dwell time is
determined by trigger event spacing.