Revision:1.0(2023-01-12)
409
• Troubleshooting suggestion
1. If the drive cannot be identified, recover the equipment.
2. If the test result is Error, restart the device and re-perform the self-test. If the problem persists,
restore the device.
3. If the test result is FAIL, replace the PC module and then the main board.
Z0102 CPU Temperature Test
• Upper-level test item
None
• Test Items
With the test software accelerating the workload, monitor whether the CPU temperature can
meet the standard in 5 minutes. Tap End Test to finish the test in the process.
Diagnostic Ultrasound System
Service Manual
9 Appendix