96
Procedures
Fine adjustment on zero at analog output.
Press the key. VADJST mode can be entered,
then load display will show “A−ZERO”.
By pressing the key again, the load display
begins flashing the present load and outputs the
analog output together. Adjust the deviation of
analog output against load with the right keys.
3
By pressing the key continuously,
continuous increase/decrease will be
provided.
The display will not vary.
: Increases the analog output value.
: Decreases the analog output value.
・ Fine adjustment on span at analog output.
Press the key.
The load display will show “A−SPAN”.
By pressing the key again, load display begins
flashing the present load and outputs the analog
output together.
Adjust the difference of analog output against
load with the right keys.
4
By pressing the key continuously,
continuous increase/decrease will be
provided.
The display will not vary.
: Increases the analog output value.
: Decreases the analog output value.