Page #147 background image
134
Display
Fuse is broken.
Replace fuse by
referring to the
paragraph 13−1.
Display doesn’t vary.
Apply re−calibration according to
section 5. Calibration.
Same condition
Start measurement
External input or
HOLD of BCD output
is used.
Inform Minebea about the contents of
failure and situation at site in details.
OL display
OL display
6
7
NO
YES
NO
NO
NO
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
YES
Make the HOLD
input open.
When HOLD is
shorted at the
time of feeding
power, no display.
Display fluctuates
abnormally.
Displays optional
values.
8
7