147
Baud rate is
set correctly.
Format for
command/data is
corrects.
Flow control
has specified
at Host.
14
Connection is correct.
By referring to the paragraph 9−4, connect
with Host properly.
By referring to the paragraph 9−4, apply
command/data correctly.
Set the following items properly according to the
specifications for HOST.
・ FUNC−30 Selection of Baud rate.
・ FUNC−31 Setting data bit length and parity.
・ FUNC−32 Setting stop bit
(Refer to the section 8. Function mode and the
paragraph 9−4.)
Change Yes into No for the flow control at
Host.
Intended values
are not transmitted in
Stream mode.
Select the output target in the Stream mode.
・ FUNC−28 Selection of output target in
Stream Mode.
(Refer to the section 8. Function mode and
the paragraph 9−4.)
By referring to the section 8. Function Mode
and the paragraph 9−4. set the ID number
correctly.
Setting ID No.
is correct.
Inform Minebea about the contents of
failure and situation at site in details.
NO
YES
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES