14
※1 When tension is applied with the application of tension type of
universal(compression/tension) type load cell, and display of “+”
direction is required, connect “green” with Terminal No. 22, and
“blue” with Terminal No. 21. As there is a case which standard
wiring color is different, please confirm the inspection data sheet of
the load cell being used.
※2 When the total length of CAB−502 specified more than 30 m, the
accuracy may be out of warranty because the resistance of cable
makes decrease the input voltage of the instrument.
※3 There is no guarantee that the function of CAB−501 will work well
if its length is totally more than 100 m.
4
Connection with 2 to 4 points of load cells and Summing type junction box (SB−310) and
CSD−814B.
i) In case of application of CAB−502(4−cores cable)
.
CAB−502
Attached short bar
(Length of CAB−502 should be within 30 m totally)
※2
※2※2
※2
Shield
※
※
※
※1 GRN
WHT
RED
※
※
※
※2 BLU
SB−310 Internal terminal connection
To CSD−814B
Suitable crimp−type terminal lugs
・Y type
RD
Short bar
Output
Point
Input
point 1
Input
point 2
Input
Point 4
or
WH
YE
YE
YEYE
YE
GR
BL
GN
GNGN
GN BL
BLBL
BLYE
YEYE
YE
YE
YEYE
YE
YE
YEYE
YE
BL
BLBL
BL
GN
GNGN
GN
WH
WHWH
WH
WH
WHWH
WH
RD
GN
GNGN
GN
BL
BLBL
BL
BL
BLBL
BLGN
GNGN
GN
WH
WHWH
WH
φ
φ
φ
φA
φ
φ
φ
φB
RD
RD
RD
Input
point 3
A:3.2 mm or less
B:7 mm or less
・Round type
B
A
WH
WHWH
WH