77
・ FUNC−06 Setting key−lock
Initial value = 000000 10
0
digit : key
10
1
digit : key
10
2
digit : key
10
3
digit : key
10
4
digit : key
10
5
digit : key
※ Each ,,,, and key can be key−locked individually.
Key−lock cancellation is made by setting “0”, and key−lock is made by setting “1”.
Moreover,during the key is locked by pressing the key for more than three
seconds with the pressed together, the key−lock will be cancelled only once
directly after that.
・ FUNC−07 Setting data width of zero tracking
Initial value = 000000 000000 = Zero tracking OFF
Setting range : 000000 ~ 000099
Unit : 0.5D
Data width of 49.5D at “000099” setting
※ Effective only when between 000001 ~ 000099 is set with the FUNC−08.
・ FUNC−08 Setting time width of zero tracking
Initial value = 000020 000000 = Zero tracking OFF
Setting range : 000000 ~ 000099
Unit : 0.1 second
Time width of 9.9 second at the setting of “000099”
※ Effective only when between 000001 ~ 000099 is set with the FUNC−07.
・ FUNC−09 Setting digital filter for stabilized filter
Initial value = 000002 000000 = Stabilized filter OFF
Setting range : 000000 ~ 000007
mean value of 2
(m+1)
times
※ Selects average times of digital filter for stabilized filter. When the figure grows
larger, the filter will be stronger, then effects from vibration and so on are scarcely
shown on the display.
※ Effective only when setting is made between 000001 ~ 000099 with the FUNC−10
and also setting is made between 000001 ~ 000999 with the FUNC−97.