・ FUNC−10 Setting time width for stabilized filter
Initial value = 000020 000000 = Stabilized filter OFF
Setting range : 000000 ~ 000999
Unit : 0.01 s
Time width of 9.99 s
at “000999” setting.
※ Effective only when setting is made between 000001 ~ 000007, with the FUNC−09
and also between 000001 ~ 000999 with the FUNC−97.
・ FUNC−11 Setting the target for comparator
Initial value = 002222 0 = Display
1 = Net weight
2 = Gross weight
10
0
digit = Comparator S1
10
1
digit = Comparator S2
10
2
digit = Comparator S3
10
3
digit = Comparator S4
※ Each comparator S1, S2, S3, and S4 can be set individually.
※ When “Display” is selected, “Load display value” becomes the target for comparator.
And when “Net weight” is selected, “Net weight” becomes the target for comparator
at the time of Tare is executed, however, “Gross weight” becomes the target at the
time of cancellation of Tare is executed. When “Gross weight” is selected, “Gross
weight” becomes the target for comparator, whichever Tare is executed or not.
・ FUNC−12 Setting direction for comparator.
Initial value = 000000 0 = Over
1 = Under
10
0
digit = Comparator S1
10
1
digit = Comparator S2
10
2
digit = Comparator S3
10
3
digit = Comparator S4
・ FUNC−13 Application condition for comparator S0
Initial value = 000000 000000 = Always OFF
000001 = ON at both S1 and S2 contact
outputs are OFF
000002 = ON at both S1 and S3 contact
outputs are OFF
000003 = ON at both S1 and S4 contact
outputs are OFF
000004 = ON at both S2 and S3 contact
outputs are OFF
000005 = ON at both S2 and S4 contact
outputs are OFF
000006 = ON at both S3 and S4 contact
outputs are OFF
000007 = ON at (Gross weight)
≧ (Maximum display value)