MOOR INSTRUMENTS moorLDI2 RESEARCH USER MANUAL
129
frame grabber board or an IEEE1394 interface card is required when using a moorLDI2-BI with a
IEEE1394 camera (this card is not required if the PC supplied has an IEEE1394 port).
* If the PC does not have an RS232 port an RS232 to USB converter is needed as the serial output
from the LDI control unit is via RS232. A suitable converter can be supplied by Moor Instruments.
Operating system – The moorLDI2-BI software can be used with Windows
XP (Home or
Professional), Windows
2000 and Windows
Vista. Moor Instruments does not support its use
with other Windows operating systems. When the software is used with Windows
Vista, it is highly
recommended that the Vista UAC (Use User Account Control) is disabled although not essential. If
UAC has to be switched on, please keep it on whenever the moorBDA software is used; otherwise
data loss may occur.
Analogue Camera Model: A half length PCI card slot is required for the CCD video camera
framegrabber board.
Firewire Camera Model: A half length PCI card slot or a PCMCIA slot is required for the CCD video
camera Firewire interface card.
* If the PC does not have an RS232 port an RS232 to USB converter is needed, as the serial output
from the LDI control unit is via RS232. A suitable converter can be supplied by Moor Instruments.
Operating system - Windows
2000 and XP or higher.
ELECTRICAL SAFETY CLASSIFICATION
Type of protection against electric shock - Class I.
Degree of protection against electric shock - Type B.
Degree of protection against ingress of liquid - IPXO (not protected).
Degree of protection against flammable anaesthetics - equipment not suitable for use in the presence
of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide.
Mode of operation - continuous.
GENERAL
Universal voltage switch mode power supply.
Range 100 to 230V AC 50VA, 50 to 60Hz.
Control Unit: Dimensions W H D mm 305 x 115 x 260: Weight 4.5kgs.
Scan Head: Dimensions W H D mm 426 x 244 x 300: Weight 8kgs.
Operating environment: Indoor hospital or laboratory use.
Operating temperature: 15-30
o
C.
STORAGE AND TRANSPORTATION CONDITIONS
Temperature: 0-45
o
C.
Humidity: 0-80% RH.
Atmospheric Pressure: 50,0 kPa – 106,0 kPa.