MOOR INSTRUMENTS moorLDI2 RESEARCH USER MANUAL
139
2.11 If measured values are outside acceptable limits (above right), then proceed to calibration with the
CALIBRATE key.
Do not calibrate if there is any doubt about the set up and accurate positioning of the calibration block. Exit
and repeat, from 2.1.
2.12 If there is excessive vibration during calibration then the Noisy Signal Message appears (below left).
Attempt calibration a second time by clicking OK, then CALIBRATE.
If the moorLDI persistently fails to calibrate, consult Moor Instruments or your authorised distributor
for further guidance.
2.13 If the ‘Calibration failed’ message appears (above right), press EXIT and re-load the calibration
utility. Pay particular attention to measurement conditions, following instructions from 2.1. If the
moorLDI persistently fails to calibrate, consult Moor Instruments or your authorised distributor for
further guidance.