Chapter E - C-ARM AND IMAGING ARM
Planmeca ProMax E-5
PANORAMIC BEAM AND PATIENT POSITIONING
MECHANISM ADJUSTMENT
Technical Manual
2 PANORAMIC BEAM AND PATIENT POSITIONING
MECHANISM ADJUSTMENT
Protect yourself from radiation when you are checking the beam alignment.
Perform the adjustments in the order given in this section.
WARNING
It is very important that the room in which the x-ray is installed and
the position from which the user operates the equipment are
correctly shielded. Since radiation safety requirements vary from
country to country and state to state it is the responsibility of the
installer to ensure that all local safety regulations are met.

Table of Contents

Other manuals for Planmeca ProMax

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProMax and is the answer not in the manual?

Planmeca ProMax Specifications

General IconGeneral
CategoryMedical Equipment
TypeCone Beam Computed Tomography (CBCT) System
SoftwarePlanmeca Romexis
Imaging ModesCBCT
Imaging Modalities2D Panoramic
X-ray GeneratorHigh Frequency Generator
Focal Spot Size0.5 mm
X-ray Tube Voltage60-90 kV
Exposure TimeVaries depending on imaging mode and parameters
DimensionsVaries significantly depending on model and configuration
WeightVaries significantly depending on model and configuration
Power Requirements50/60 Hz
Patient Positioningseated

Related product manuals