Chapter E - C-ARM AND IMAGING ARM
Planmeca ProMax E-103
PROMAX 3D: UPDATING SENSOR SOFTWARE
Technical Manual
12 PROMAX 3D: UPDATING SENSOR SOFTWARE
NOTE The sensor software is updated from an external PC.
a) Start the ProMax3D Tool program.
b) Select Software Update Sensor Software.
Figure 139
c) Select the software from the browser (e.g. 3D_sensor_2.2.0.0.R.EthUpgrd.ldr).
d) Wait until the software has been loaded successfully, “Sensor Software Updated” message
will be displayed.
e) Select Software Sensor Software Version to check that the software was updated.