Chapter B - INFORMATION DISPLAYS
B-28 Planmeca ProMax
SERVICE SETTINGS
Technical Manual
To set the cephalometric exposure values select Set ceph parameters limits (i54.2). The
display shown below appears. Select the maximum and minimum kilovolt, and minimum
milliampere values with the arrow fields and touch the Done field.
Figure 41
To set the 3D exposure values select Set 3D parameters limits (i54.3). The display shown
below appears. Select the maximum and minimum kilovolt, and minimum milliampere
values with the arrow fields and touch the Done field.
Figure 42
5.5 Factory default settings
The Factory default settings (i550) will turn up only in case the memory card (RTC) has lost
its configuration and user settings or it has been replaced to a new PCB. In this case the
procedure should be done before using the X-ray unit.

Table of Contents

Other manuals for Planmeca ProMax

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProMax and is the answer not in the manual?

Planmeca ProMax Specifications

General IconGeneral
CategoryMedical Equipment
TypeCone Beam Computed Tomography (CBCT) System
SoftwarePlanmeca Romexis
Imaging ModesCBCT
Imaging Modalities2D Panoramic
X-ray GeneratorHigh Frequency Generator
Focal Spot Size0.5 mm
X-ray Tube Voltage60-90 kV
Exposure TimeVaries depending on imaging mode and parameters
DimensionsVaries significantly depending on model and configuration
WeightVaries significantly depending on model and configuration
Power Requirements50/60 Hz
Patient Positioningseated

Related product manuals