Chapter G - COLUMN
Planmeca ProMax G-11
REPLACING THE COLUMN MOTOR
Technical Manual
3 REPLACING THE COLUMN MOTOR
3.1 Replacing the motor
WARNING
Care must be taken when replacing the column motor. Removing
par ts of the lifting assembly may enable the column to fall causing
a dangerous situation and/or damage to the equipment. The down-
ward movement of the telescopic column must be eliminated
before detaching the parts.
WARNING
The X-ray unit contains live mains voltage parts. Always unplug
the unit from mains outlet before attempting to service the Input
module .
CAUTION At least two persons are needed when replacing the column motor.
Never try to replace the motor on your own.

Table of Contents

Other manuals for Planmeca ProMax

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProMax and is the answer not in the manual?

Planmeca ProMax Specifications

General IconGeneral
CategoryMedical Equipment
TypeCone Beam Computed Tomography (CBCT) System
SoftwarePlanmeca Romexis
Imaging ModesCBCT
Imaging Modalities2D Panoramic
X-ray GeneratorHigh Frequency Generator
Focal Spot Size0.5 mm
X-ray Tube Voltage60-90 kV
Exposure TimeVaries depending on imaging mode and parameters
DimensionsVaries significantly depending on model and configuration
WeightVaries significantly depending on model and configuration
Power Requirements50/60 Hz
Patient Positioningseated

Related product manuals