Chapter A - GENERAL & TECHNICAL DATA
A-20 Planmeca ProMax
EMC INFORMATION
Technical Manual
9 EMC INFORMATION
WARNING
Use of any accessories and cables other than those specified in Planmeca Pro-
Max X-ray unit’s documentation, with the exception of cables sold by Planmeca
as replacement parts for internal components, may result in increased emis-
sion or decreased immunity of the X-ray unit.
WARNING
Planmeca ProMax X-ray unit should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, Planmeca ProMax X-
ray unit should be observed to verify its normal operation in the configuration
in which it is used.
Guidance and manufacturer’s declaration - electromagnetic emissions
Planmeca ProMax X-ray unit is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of the Planmeca ProMax X-ray unit should assure that
it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
Planmeca ProMax X-ray unit uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Planmeca ProMax X-ray unit is suitable for
use in all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply net-
work that supplies buildings used for domestic
purposes.
Harmonic emis-
sions
IEC 61000-3-2
Class A
Voltage fluctua-
tions/flicker emis-
sions
IEC 61000-3-3
Complies