Chapter A - GENERAL & TECHNICAL DATA
Planmeca ProMax A-21
EMC INFORMATION
Technical Manual
Guidance and manufacturer’s declaration - electromagnetic immunity
Planmeca ProMax X-ray unit is intended for use in the electromagnetic environment specified below. The
customer or the user of Planmeca ProMax X-ray unit should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are cov-
ered with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to earth
±1 kV line to line
±2 kV line to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for5s
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of Planmeca ProMax X-ray unit
requires continued operation dur-
ing power mains interruptions, it
is recommended that Planmeca
ProMax X-ray unit be powered
from an uninterruptible power
supply.
Power frequency(
50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment. The power frequency mag-
netic field should be measured in
the intended installation location
to assure that it is sufficiently low.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.

Table of Contents

Other manuals for Planmeca ProMax

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProMax and is the answer not in the manual?

Planmeca ProMax Specifications

General IconGeneral
CategoryMedical Equipment
TypeCone Beam Computed Tomography (CBCT) System
SoftwarePlanmeca Romexis
Imaging ModesCBCT
Imaging Modalities2D Panoramic
X-ray GeneratorHigh Frequency Generator
Focal Spot Size0.5 mm
X-ray Tube Voltage60-90 kV
Exposure TimeVaries depending on imaging mode and parameters
DimensionsVaries significantly depending on model and configuration
WeightVaries significantly depending on model and configuration
Power Requirements50/60 Hz
Patient Positioningseated