EasyManuals Logo

Planmeca ProMax Technical Manual

Planmeca ProMax
378 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #160 background imageLoading...
Page #160 background image
Chapter E - C-ARM AND IMAGING ARM
E-62 Planmeca ProMax
PROMAX 3D: ADJUSTMENTS & CALIBRATIONS
Technical Manual
3.3 Beam adjustment
Select the standard 3D exposure program on the control panel of the X-ray unit.
To be able to check the radiation beam position without the C-arm moving enter the 3D
collimator calibration mode as explained below.
Beam vertical position
If the X-ray beam is too low or too high, the vertical position of the beam must be adjusted.
Select Primary collimator calibration (i610) from the display that appears when you select
Technical calibrations.
Depending on the sensor you are using, select 3D or 3D s mode Y-collimator calibration
(i61.11 or i61.13) from the list.
The position of the X-ray beam can now be checked without the C-arm moving.
Figure 79
The corresponding Y-collimator calibration (i61.11 or i61.13) display will appear. The vertical
position of the X-ray beam is adjusted in this mode.
Remove the sensor head from the quick connector mechanism. Attach the sensor head
alignment tool to the connector as shown in Fig. 80.
Manually position the C-arm to a convenient position for viewing the sensor head alignment
tool.
X-ray units
with 3D sensor
X-ray units
with 3D s sensor
(V1 or V2)

Table of Contents

Other manuals for Planmeca ProMax

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProMax and is the answer not in the manual?

Planmeca ProMax Specifications

General IconGeneral
CategoryMedical Equipment
TypeCone Beam Computed Tomography (CBCT) System
SoftwarePlanmeca Romexis
Imaging ModesCBCT
Imaging Modalities2D Panoramic
X-ray GeneratorHigh Frequency Generator
Focal Spot Size0.5 mm
X-ray Tube Voltage60-90 kV
Exposure TimeVaries depending on imaging mode and parameters
DimensionsVaries significantly depending on model and configuration
WeightVaries significantly depending on model and configuration
Power Requirements50/60 Hz
Patient Positioningseated

Related product manuals