Chapter G - COLUMN
G-4 Planmeca ProMax
ADJUSTMENTS
Technical Manual
1.2 Calibrating the column motor position sensor
CAUTION Protect yourself against electrical shock. The unit contains live parts on
some PCBs and connectors.
Unscrew the four attachment screws of the stationary column top cover with the 3mm Allen
key and remove the cover (Fig. 21, 1).
Using the height adjusting buttons, drive the telescopic column into position (Fig. 21, 2) where
the reference mark on the telescopic column and the upper edge of the stationary column are
aligned (Fig. 21, 3).
Figure 21
1
2
3
Digi_tele.eps

Table of Contents

Other manuals for Planmeca ProMax

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProMax and is the answer not in the manual?

Planmeca ProMax Specifications

General IconGeneral
CategoryMedical Equipment
TypeCone Beam Computed Tomography (CBCT) System
SoftwarePlanmeca Romexis
Imaging ModesCBCT
Imaging Modalities2D Panoramic
X-ray GeneratorHigh Frequency Generator
Focal Spot Size0.5 mm
X-ray Tube Voltage60-90 kV
Exposure TimeVaries depending on imaging mode and parameters
DimensionsVaries significantly depending on model and configuration
WeightVaries significantly depending on model and configuration
Power Requirements50/60 Hz
Patient Positioningseated

Related product manuals