Chapter A - GENERAL & TECHNICAL DATA
A-4 Planmeca ProMax
TECHNICAL SPECIFICATIONS
Technical Manual
SmartPan image properties:
Flat panel pixel size 127 m
Flat panel active surface 0.8 x 13 cm (0.31 x 5.1 in.)
Line voltage 100 - 132 V~ / 50- 60 Hz
180 - 240 V~ / 50 Hz
Line current 8 - 15 A
Line harmonics cos better than 0.9
Max. permissible apparent
impedance of supply mains 0.5 (100VAC)
Maximum continuous
heat dissipation < 250W
Electrical classification Class I, type B
Fuses
Weight 119 kg (263 lbs)
137 kg (304 lbs) with Cephalostat
Colour RAL 9016
Environmental requirements
Ambient temperature Pan / ceph X-ray unit:
Operating +5°C to +40°C
Storage ±0°C to +50°C
3D / 3D s X-ray unit:
Operating +10°C to +30°C
Storage ±0°C to +50°C
Humidity 15% - 85%
Original manufacturer
PLANMECA Oy, Asentajankatu 6, FIN-00880 Helsinki, FINLAND
phone: +358 20 7795 500, fax: +358 20 7795 555, www.planmeca.com
REPLACEABLE FUSES
180-240V~ 100-132V~
TYPE
2 pieces (user replaceable) 8A FF 16A FF /500V 195100 ELU

Table of Contents

Other manuals for Planmeca ProMax

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProMax and is the answer not in the manual?

Planmeca ProMax Specifications

General IconGeneral
CategoryMedical Equipment
TypeCone Beam Computed Tomography (CBCT) System
SoftwarePlanmeca Romexis
Imaging ModesCBCT
Imaging Modalities2D Panoramic
X-ray GeneratorHigh Frequency Generator
Focal Spot Size0.5 mm
X-ray Tube Voltage60-90 kV
Exposure TimeVaries depending on imaging mode and parameters
DimensionsVaries significantly depending on model and configuration
WeightVaries significantly depending on model and configuration
Power Requirements50/60 Hz
Patient Positioningseated