Page 9-8 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
DEFAULTS (cont.)
The factory-set default Control settings are:
• Some Controls are not applicable to all LTV
®
ventilators; see Chapter 6 - Controls, for detailed
information concerning specific Controls.
Control
- Default Control - Default
Breath Rate - 12 bpm Low Pressure O
2
Source - Off
Control Lock - On Pressure Control - 1 cmH
2
O
Data Display Scrolling - Auto-On Pressure Support - 1 cmH
2
O
High Pressure Limit - 20 cmH
2
O Sensitivity - 2 Lpm
Inspiratory/Expiratory Hold - Off Tidal Volume - 500 ml
Inspiratory Time - 1.5 sec Ventilation Mode - Assist / Control
Low Minute Volume - 2.5 Lpm Volume / Pressure Mode - Volume
Low Pressure - 5 cmH
2
O O
2
% (O
2
Flush) - 21%
The factory-set default Extended Features settings are:
• Some Extended Features are not applicable to all LTV
®
ventilators; see Chapter 10 - Extended
Features, for detailed information concerning specific Features.
Feature
- Default
Feature - Default
O
2
Flush Period - 3 min Leak Compensation - On
Alarm Volume - 85 dBA LPP Alarm - All Breaths
Apnea Interval - 20 sec NPPV Mode - Off
Com Setting
39
- Data
O
2
Duration
Cylinder Press - 2000 psi or
138 bar
Control Unlock -
Easy
O
2
Duration
Cylinder Size -
622 liters
Date Format
39
- mm/dd/yyyy PC Flow Termination - Off
High f Alarm - High f Off PIP LED - On
High f Alarm Delay - 30 sec PNT Assist - Normal
High PEEP Alarm - High PEEP Off Rise Time Profile - 4
HP Alarm Delay - No Delay Variable Flow Termination - 25%
Language
39
- English Variable Time Termination - 1.5 sec
39
This feature is not reset to default values when the SET DEFAULTS option is used in Extended Features.