Page 10-18 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Set Time
Use the Set Time option to view or set the current time stored in the ventilator.
To view the Time:
1) Push the Select button while SET TIME is displayed.
2) The current time is displayed.
3) Push the Control Lock button to exit.
To modify the Time:
1) Push the Select button while SET TIME is displayed.
2) The current date is displayed as hh:mm:ss.
3) Push the Select button, HOUR xx is displayed.
4) Turn the Set Value knob until the desired hour is displayed.
5) Push the Select button, MIN xx is displayed.
6) Turn the Set Value knob until the desired minute is displayed.
7) Push the Select button to accept the new date. The seconds are automatically reset to 00.
Range: 00:00:00 - 23:59:59
Date Format
Use the Date Format option to select the display format for the current date.
To modify the Date Format:
1) Push the Select button while DATE FORMAT is displayed.
2) MM/DD/YYYY or the currently selected date format is displayed.
3) Turn the Set Value knob until the desired format is displayed.
4) Push the Select button.
Options: MM/DD/YYYY, DD/MM/YYYY, YYYY/MM/DD