Operator’s Manual LTV
®
Series ventilator Page 6-11
p/n 10664, Rev. Y
Low Minute Volume
The Low Minute Volume alarm sets the minimum expected exhaled Minute Volume. The exhaled
Minute Volume is recalculated after every breath. If the Minute Volume does not meet or exceed the
Low Minute Volume setting:
• A LOW MIN VOL alarm is displayed
• The audible alarm is sounded
The Low Minute Volume alarm is not active in NPPV mode.
To set the Low Minute Volume alarm:
1) Push the Low Minute Volume button.
2) Change the setting using the Set Value knob.
Range: Off, 0.1 - 99 L
WARNING
Low Minute Volume Control Settings - The Low Minute Volume control should be set to its highest
clinically appropriate value. If there is a clinical need to set the Low Minute Volume alarm to lower
values or off (“- - -“), perform a clinical assessment to determine if an alternative monitor (i.e. a Pulse
Oxymeter with an audible alarm, or a Cardio Respiratory Monitor) should be used.
AVERTISSEMENT
Réglages du contrôle de volume bas par minute - Le contrôle du volume bas par minute doit être
ajusté à la plus haute valeur clinique appropriée. Si l’alarme de volume bas par minute doit être
ajustée à des valeurs inférieures ou mise à l’arrêt (“- - -“) pour satisfaire aux besoins cliniques,
effectuer une évaluation clinique afin de déterminer si l’utilisation d’un autre moniteur (c.-à-d.,
sphygmo-oxymètre muni d’une alarme sonore ou un moniteur cardio-respiratoire) s’avère pertinente.