Operator’s Manual LTV
®
Series ventilator Page 9-13
p/n 10664, Rev. Y
HIGH PEEP
When the patient circuit positive end expiratory pressure (PEEP) exceeds the High PEEP alarm
setting, the HIGH PEEP alarm is generated.
When a HIGH PEEP alarm occurs:
The HIGH PEEP message is flashed in the display window.
The audible alarm is sounded.
To reset the HIGH PEEP alarm:
1) Push the Silence / Reset button once to silence the audible alarm.
2) Push the Silence / Reset button twice
to reset the alarm (silences the audible alarm and clears
the flashing display).